Kinematic alignment in robotic total knee arthroplasty

https://doi.org/10.18019/1028-4427-2024-30-6-845-854

EDN: VQQOFA

Abstract

Introduction There are two main concepts of total knee arthroplasty: mechanical and anatomical alignment of the lower limb axis. Howell et al. (2013) proposed the concept of kinematic alignment, the main idea of which is to preserve the level of the joint line and the axis of the lower limb that patients had before the onset of osteoarthritis. Initially, kinematic alignment was proposed to be performed with individual guides based on the results of CT/MRI scans but they took a long time to manufacture, were difficult to install, broke down, and were quite expensive. Introduction of robotic orthopaedic systems into clinical practice enabled to plan and perform bone resection with high accuracy, to install the components of the implant system according to  the  necessary concept thus providing new opportunities for the application of kinematic alignment, which was the purpose of our study.

Purpose To study the possibilities of a robotic surgical system in performing restricted kinematic alignment in total knee arthroplasty (TKA).

Materials and methods A prospective single-center study was conducted in 47 patients (12 men and 35 women) with knee osteoarthritis in Kellgren – Lawrence grades 3–4, an average age of (65.87 ± 7.4) years, an average BMI of (31.3 ± 4.7) kd/m2 , median HKA angle of 175°, median LDFA of 87°, median MPTA of 87°. The patients underwent robotic total knee arthroplasty (RoTKA) with the method of restricted kinematic alignment.

Results On the control whole-leg radiographs, the average HKA angle after surgery was (176 ± 1.5)°. In 42.6 % of cases, the deviation from the plan was within ± 1°, the deviation ± 2° in 44.7 % of cases, and in the remaining 12.7 % of cases the deviation was negative.

Discussion In the literature, we did not find the results of radiographic evaluation of the HKA angle in the coronal view before and after robotic total knee arthroplasty and their comparison with the results of preoperative planning using the kinematic alignment method of the limb axis. The results we obtained show high accuracy of the implementation of the preoperative plan.

Conclusion A personalized approach to TKA with application of an autonomous robotic system effectively provides kinematic alignment of the axis of the lower limb with an accuracy of up to 2° in 87.3 % of patients.

About the Authors

A. V. Lychagin
First Moscow State Medical University named after I.M. Sechenov (Sechenov University)
Russian Federation

Alexey V. Lychagin — Doctor of Medical Sciences,  Professor, Director of the Clinic, Head of the Department

Moscow



A. A. Gritsyuk
First Moscow State Medical University named after I.M. Sechenov (Sechenov University)
Russian Federation

Andrey A. Gritsyuk — Doctor of Medical Sciences, Professor of the Department

Moscow



E. B. Kalinsky
First Moscow State Medical University named after I.M. Sechenov (Sechenov University)
Russian Federation

Eugene B. Kalinsky — Doctor of Medical Sciences, Professor of the Department

Moscow



Ya. A. Rukin
First Moscow State Medical University named after I.M. Sechenov (Sechenov University)
Russian Federation

Yaroslav A. Rukin — Candidate of Medical Sciences, Head of the Department, Assistant Professor

Moscow



M. P. Elizarov
First Moscow State Medical University named after I.M. Sechenov (Sechenov University)
Russian Federation

Mikhail P. Elizarov — Assistant of the Department

Moscow



A. A. Gritsyuk
First Moscow State Medical University named after I.M. Sechenov (Sechenov University)
Russian Federation

Andrey А. Gritsyuk — post-graduate student

Moscow



M. Ya. Gavlovsky
First Moscow State Medical University named after I.M. Sechenov (Sechenov University)
Russian Federation

Maxim Ya. Gavlovsky — post-graduate student

Moscow



K. Kh. Tomboidi
First Moscow State Medical University named after I.M. Sechenov (Sechenov University)
Russian Federation

Konstantin Kh. Tomboidi — clinical resident

Moscow



M. L. Babaeva
First Moscow State Medical University named after I.M. Sechenov (Sechenov University)
Russian Federation

Milana L. Babaeva — clinical resident

Moscow



References

1. Bellemans J. Neutral mechanical alignment: a requirement for successful TKA: opposes. Orthopedics. 2011;34(9):e507-e509. doi: 10.3928/01477447-20110714-41

2. Eckhoff DG, Bach JM, Spitzer VM, et al. Three-dimensional mechanics, kinematics, and morphology of the knee viewed in virtual reality. J Bone Joint Surg Am. 2005;87 Suppl 2:71-80. doi: 10.2106/JBJS.E.00440

3. Almaawi AM, Hutt JRB, Masse V, et al. The impact of mechanical and restricted kinematic alignment on knee anatomy in total knee arthroplasty. J Arthroplasty. 2017;32(7):2133-2140. doi: 10.1016/j.arth.2017.02.028

4. Lin YH, Chang FS, Chen KH, et al. Mismatch between femur and tibia coronal alignment in the knee joint: classification of five lower limb types according to femoral and tibial mechanical alignment. BMC Musculoskelet Disord. 2018;19(1):411. doi: 10.1186/s12891-018-2335-9

5. MacDessi SJ, Griffiths-Jones W, Harris IA, et al. Coronal Plane Alignment of the Knee (CPAK) classification. Bone Joint J. 2021;103-B(2):329-337. doi: 10.1302/0301-620X.103B2.BJJ-2020-1050.R1

6. Hirschmann MT, Hess S, Behrend H, et al. Phenotyping of hip-knee-ankle angle in young non-osteoarthritic knees provides better understanding of native alignment variability. Knee Surg Sports Traumatol Arthrosc. 2019;27(5):13781384. doi: 10.1007/s00167-019-05507-1

7. Robinson RP. The early innovators of today's resurfacing condylar knees. J Arthroplasty. 2005;20(1 Suppl 1):2-26. doi: 10.1016/j.arth.2004.11.002

8. Insall JN, Binazzi R, Soudry M, Mestriner LA. Total knee arthroplasty. Clin Orthop Relat Res. 1985;(192):1322.

9. Matassi F, Pettinari F, Frasconà F et al. Coronal alignment in total knee arthroplasty: a review. J Orthop Traumatol. 2023;24(1):24. doi: 10.1186/s10195-023-00702-w

10. Begum FA, Kayani B, Magan AA, et al. Current concepts in total knee arthroplasty : mechanical, kinematic, anatomical, and functional alignment. Bone Jt Open. 2021;2(6):397-404. doi: 10.1302/2633-1462.26.BJO2020-0162.R1

11. Hungerford DS, Krackow KA. Total joint arthroplasty of the knee. Clin Orthop Relat Res. 1985;(192):23-33.

12. Rivière C, Iranpour F, Auvinet E, et al. Alignment options for total knee arthroplasty: A systematic review. Orthop Traumatol Surg Res. 2017;103(7):1047-1056. doi: 10.1016/j.otsr.2017.07.010

13. McClelland JA, Webster KE, Feller JA. Gait analysis of patients following total knee replacement: a systematic review. Knee. 2007;14(4):253-263. doi: 10.1016/j.knee.2007.04.003

14. Bellemans J, Colyn W, Vandenneucker H, Victor J. The Chitranjan Ranawat award: is neutral mechanical alignment normal for all patients? The concept of constitutional varus. Clin Orthop Relat Res. 2012;470(1):45-53. doi: 10.1007/s11999-011-1936-5

15. Howell SM, Papadopoulos S, Kuznik KT, Hull ML. Accurate alignment and high function after kinematically aligned TKA performed with generic instruments. Knee Surg Sports Traumatol Arthrosc. 2013;21(10):22712280. doi: 10.1007/s00167-013-2621-x

16. Calliess T, Bauer K, Stukenborg-Colsman C, et al. PSI kinematic versus non-PSI mechanical alignment in total knee arthroplasty: a prospective, randomized study. Knee Surg Sports Traumatol Arthrosc. 2017;25(6):1743-1748. doi: 10.1007/s00167-016-4136-8

17. Blakeney WG, Vendittoli PA. Kinematic Alignment Total Knee Replacement with Personalized Instruments. 2020. In: Rivière C, Vendittoli PA, eds. Personalized Hip and Knee Joint Replacement [Internet]. Cham (CH): Springer; 2020. Chapter 25

18. Morcos MW, Uhuebor D, Vendittoli PA. Overview of the different personalized total knee arthroplasty with robotic assistance, how choosing? Front Surg. 2023;10:1120908. doi: 10.3389/fsurg.2023.1120908

19. Wan X, Su Q, Wang D, et al. Robotic arm-assisted total knee arthroplasty improves preoperative planning and intraoperative decision-making. J Orthop Surg Res. 2021;16(1):670. doi: 10.1186/s13018-021-02815-6

20. Lychagin AV, Gritsyuk AA, Rukin YA, Elizarov MP. The history of the development of robotics in surgery and orthopedics (Literature review). Department of Traumatology and Orthopedics. 2020;(1):13-19. doi: 10.17238/issn2226-2016.2020.1.13-19

21. Lychagin AV, Gritsyuk AA, Rukin YaA, et al. Clinical evaluation and accuracy of mechanical axis alignment in robotic total knee arthroplasty. Genij Ortopedii. 2023;29(5):487-494. doi: 10.18019/1028-4427-2023-29-5-487-494

22. Balaguer-Castro M, Torner P, Jornet-Gibert M, Martínez-Pastor JC. Current situation of robotics in knee prosthetic surgery, a technology that has come to stay? Rev Esp Cir Ortop Traumatol. 2023;67(4):334-341. doi: 10.1016/j.recot.2022.10.011

23. Lychagin AV, Gritsyuk AA, Elizarov MP et al. Randomized double controlled study of the accuracy of lower limb mechanical axis alignment. Medical Bulletin of the Main Military Clinical Hospital named after N.N. Burdenko. 2023;4(3):40-47. doi: 10.53652/2782-1730-2023-4-3-40-47

24. Massé V, Cholewa J, Shahin M. Personalized alignment™ for total knee arthroplasty using the ROSA® Knee and Persona® knee systems: Surgical technique. Front Surg. 2023;9:1098504. doi: 10.3389/fsurg.2022.1098504

25. Liu B, Feng C, Tu C. Kinematic alignment versus mechanical alignment in primary total knee arthroplasty: an updated meta-analysis of randomized controlled trials. J Orthop Surg Res. 2022;17(1):201. doi: 10.1186/s13018-022-03097-2

26. Vendittoli PA, Martinov S, Blakeney WG. Restricted kinematic alignment, the fundamentals, and clinical applications. Front Surg. 2021;8:697020. doi: 10.3389/fsurg.2021.697020

27. Ollivier B, Vandenneucker H, Vermue H, Luyckx T. A robotic-assisted simulation of kinematic alignment in TKA leads to excessive valgus and internal rotation in valgus knees. Knee Surg Sports Traumatol Arthrosc. 2023;31(11):4747-4754. doi: 10.1007/s00167-023-07504-x

28. Aflatooni JO, Wininger AE, Park KJ, Incavo SJ. Alignment options and robotics in total knee arthroplasty. Front Surg. 2023;10:1106608. doi: 10.3389/fsurg.2023.1106608

29. Morrisey ZS, Barra MF, Guirguis PG, Drinkwater CJ. Transition to robotic total knee arthroplasty with kinematic alignment is associated with a short learning curve and similar acute-period functional recoveries. Cureus. 2023;15(5):e38872. doi: 10.7759/cureus.38872

30. Kafelov M, Batailler C, Shatrov J, et al. Functional positioning principles for image-based robotic-assisted TKA achieved a higher Forgotten Joint Score at 1 year compared to conventional TKA with restricted kinematic alignment. Knee Surg Sports Traumatol Arthrosc. 2023;31(12):5591-5602. doi: 10.1007/s00167-023-07609-3

31. Huber K, Christen B, Calliess S, Calliess T. True kinematic alignment is applicable in 44% of patients applying restrictive indication criteria-A retrospective analysis of 111 TKA using robotic assistance. J Pers Med. 2021;11(7):662. doi: 10.3390/jpm11070662.


Review

For citations:


Lychagin A.V., Gritsyuk A.A., Kalinsky E.B., Rukin Ya.A., Elizarov M.P., Gritsyuk A.A., Gavlovsky M.Ya., Tomboidi K.Kh., Babaeva M.L. Kinematic alignment in robotic total knee arthroplasty. Genij Ortopedii. 2024;30(6):845-854. https://doi.org/10.18019/1028-4427-2024-30-6-845-854. EDN: VQQOFA

Views: 438


Creative Commons License
This work is licensed under a Creative Commons Attribution 4.0 License.


ISSN 1028-4427 (Print)
ISSN 2542-131X (Online)