Comparative analysis of the lower leg multiapical deformity correction by various methods of orthopedic hexapods usage
https://doi.org/10.18019/1028-4427-2025-31-3-287-296
Abstract
Introduction The technique of multiapical deformity correction with orthopaedic hexapods can be accepted as "standard." This requires several software calculations a and lengths changes in the 12–18 struts. The "spring" technique was designed to address these disadvantages.
Purpose To analyze the treatment results of patients with lower leg multiapical deformities corrected by "standard" and "spring" techniques
Methods The data of patient group 1 (standard technique, n = 17) and patient group 2 (spring technique, n = 17), were used. Correction accuracy, duration of correction, and fixation periods were compared. In patients requiring lengthening, fixation and osteosynthesis indices were additionally analyzed. Quality of life and segment function were assessed using the LEFS questionnaire.
Results There was no statistically significant difference between the groups when comparing each of the parameters studied. When comparing LEFS parameters before external fixation (EF) and 2–3 months after frame removal, a statistically significant difference was observed in patients within each of the first and second groups.
Discussion The correction accuracy in each group was 94.4 %, which, along with the absence of a significant difference in the duration of treatment and data from the LEFS questionnaire, indicates equal clinical effectiveness of both methods. It was noted that treatment with various modifications of circular external fixators is equally uncomfortable for patients. At the same time, the need for only one calculation in a hexapod software instead of 2–3, the ability to change the length of 6 struts instead of 12–18, as well as simpler assembly and less cumbersome construct if there is small (less than 10–12 cm) distance between the rings are advantages of a "spring" technique.
Conclusion Based on the analysis of the criteria used for evaluation, it can be concluded that the “spring” technique for correcting lower leg multiapical deformities is as effective as the “standard” technique. The advantages of the "spring" technique are associated with the greater convenience of its use: both for the orthopedic surgeon and the patient.
About the Authors
E. S. GolovenkinRussian Federation
Evgeniy S. Golovenkin — Clinical Postgraduate Student
St. Petersburg
L. N. Solomin
Russian Federation
Leonid N. Solomin — Doctor of Medical Sciences, Professor, Leading Researcher
St. Petersburg
References
1. Vilensky VA. Development of the fundamentals of a new technology for treating patients with diaphyseal injuries of long bones based on a transosseous apparatus with passive computer navigation properties (experimental and clinical study): Autoref. kand. dis. Saint Petersburg; 2009:31. Available at: https://viewer.rsl.ru/ru/rsl01003482854?page=8&rotate=0&theme=white. Accessed Mar 5, 2025. (In Russ.)
2. Ganger R, Radler C, Speigner B, Grill F. Correction of post-traumatic lower limb deformities using the Taylor spatial frame. Int Orthop. 2010;34(5):723-730. doi: 10.1007/s00264-009-0839-5.
3. Vilensky VA, Pozdeev AP, Bukharev EV, et al. Orthopedic hexapods: history, present and prospects. Pediatric Traumatology, Orthopaedics and Reconstructive Surgery. 2015;3(1):61-69. (In Russ.) doi: 10.17816/PTORS3161-69.
4. Solomin LN, Shchepkina EA, Korchagin KL, et al. The new method of long bone multilevel deformities correction using the orthopedic hexapod (preliminary report). Traumatology and Orthopedics of Russia. 2017;23(3):103-109. (In Russ.) doi: 10.21823/2311-2905-2017-23-3-103-109.
5. Lu Y, Li J, Qiao F, et al. Correction of severe lower extremity deformity with digital hexapod external fixator based on CT data. Eur J Med Res. 2022;27(1):252. doi: 10.1186/s40001-022-00887-6.
6. Naqui SZ, Thiryayi W, Foster A, et al. Correction of simple and complex pediatric deformities using the Taylor-Spatial Frame. J Pediatr Orthop. 2008;28(6):640-647. doi: 10.1097/BPO.0b013e3181831e99.
7. Keshet D, Eidelman M. Clinical utility of the Taylor spatial frame for limb deformities. Orthop Res Rev. 2017;9:51-61. doi: 10.2147/ORR.S113420.
8. Riganti S, Nasto LA, Mannino S, et al. Correction of complex lower limb angular deformities with or without length discrepancy in children using the TL-HEX hexapod system: comparison of clinical and radiographical results. J Pediatr Orthop B. 2019;28(3):214-220. doi: 10.1097/BPB.0000000000000573.
9. Vilensky V.A., Zakharyan E.A., Zubairov T.F. et al. Treatment of two-level deformities of lower leg bones: two hexapods or one? Modern problems of science and education. 2019;(6). (In Russ.) doi: 10.17513/spno.29352.
10. Ray V, Popkov D, Lascombes P, et al. Simultaneous multisegmental and multifocal corrections of complex lower limb deformities with a hexapod external fixator. Orthop Traumatol Surg Res. 2023;109(3):103042. doi: 10.1016/j.otsr.2021.103042.
11. Massobrio M, Mora R. Hexapod External Fixator Systems: Principles and Current Practice in Orthopaedic Surgery. Springer International Publishing; 2021:311.
12. Trombetti A, Al-Daghri N, Brandi ML et al. Interdisciplinary management of FGF23-related phosphate wasting syndromes: a Consensus Statement on the evaluation, diagnosis and care of patients with X-linked hypophosphataemia. Nat Rev Endocrinol. 2022;18(6):366-384. doi: 10.1038/s41574-022-00662-x.
13. Golovenkin ES, Solomin LN. Correction of Multiapical Deformities of Long Bones of the Lower Extremities: A Review. Traumatology and Orthopedics of Russia. 2023;29(4):134-146. doi: 10.17816/2311-2905-11174.
14. Iobst C., Ferreira N., Kold S. A Review and Comparison of Hexapod External Fixators: Current Concept Review. J Pediatr Orthop Soc North Am. 2023;5(1):627. doi: 10.55275/JPOSNA-2023-627.
15. Caton J. Bilateral lengthening of lower limbs in short subjects in France. Results of the GEOP survey; our experience: Treatment of inequalities in length of lower limbs and short subjects in children and adolescents: Symposium under the direction of J Caton (Lyon). Rev Chir Orthop. 1991;77(1):74-77. (In Fran.)
16. Vilensky VA, Pozdeev AA, Zubairov TF, et al. Treatment of pediatric patients withlower extremity deformities using software-assisted ORTHO-SUV frame: analysis of 213 cases. Pediatric Traumatology, Orthopaedics and Reconstructive Surgery. 2016;4(4):21-32. doi: 10.17816/PTORS4421-32.
17. Zakharyan EA. Complex treatment of lower limb deformities in children with congenital pseudoarthrosis of the tibia: Kand. Dis. St. Petersburg; 2017:154. Available at: http://dissovet.rniito.ru/ds2/upload/files/zaharian_ea/diss.pdf. Accessed Mar 05, 2025. (In Russ.)
18. Rokhoev S.A. Justification for the use of an orthopedic hexapod in the treatment of patients with knee joint contractures (anatomical and clinical study): Kand. Dis. St. Petersburg; 2022:181. Available at: http://dissovet.rniito.ru/ds2/upload/files/rohoev/dissert.pdf. Accessed Mar 05, 2025. (In Russ.)
19. Lebedkov IV. Comparative evaluation of the effectiveness of combined transosseous and intramedullary osteosynthesis and lengthening according to Ilizarov in restoring the length of the tibia and femur (experimental and clinical study): Kand. Dis. St. Petersburg; 2023:200. Available at: http://dissovet.rniito.ru/ds2/upload/files/lebedkov/dissert.pdf. Accessed Mar 05, 2025. (In Russ.)
20. Gavrilov DV, Solomin LN. Comparative analysis of the reduction capabilities of orthopedic hexapod Ortho-SUV frame and its minimized (pediatric) version (experimental study). Genij Ortopedii. 2023;29(3):270-276. doi: 10.18019/1028-4427-2023-29-3-270-276.
21. Eidelman M, Bialik V, Katzman A. Correction of deformities in children using the Taylor spatial frame. J Pediatr Orthop B. 2006;15(6):387-395. doi: 10.1097/01.bpb.0000228380.27239.8a.
22. Koren L, Keren Y, Eidelman M. Multiplanar Deformities Correction Using Taylor Spatial Frame in Skeletally Immature Patients. Open Orthop J. 2016;10:71-79. doi: 10.2174/1874325001610010603.
Review
For citations:
Golovenkin E.S., Solomin L.N. Comparative analysis of the lower leg multiapical deformity correction by various methods of orthopedic hexapods usage. Genij Ortopedii. 2025;31(3):287-296. https://doi.org/10.18019/1028-4427-2025-31-3-287-296